Maintenance and precautions of Muffle Furnace



1. When a muffle furnace for the first time use or long-term disabled again after use and must be oven. The time should be at room temperature to 200C furnace for four hours. 200C to 600C for four hours. When in use, shall not exceed a maximum rated temperature of the temperature to prevent burning electric components. Prohibits the perfusion fluids and soluble metal in the furnace,muffle furnace best in less than the maximum working temperature is below 50degrees, furnace has a longer life expectancy.

2. Muffle furnace and controller must be relative humidity no more than 85%,there is no conductive dusts, explosive gas or corrosive gas establishments.Any metals require heating, such as oil, a large number of volatile gas will affect the electric components surface and corrosion, damage and shorten your life. Therefore, the heating should be timely prevention and sealing containers or appropriate holes be excluded.

3. Muffle the controller should be limited to the environmenttemperature range 0-40C.

4. According to the technical requirements, regularly check the eleltric wiring for well, controllers, indicators during the pointer movement without stuck using potentiometer proof instrument due to magnetic,demagnetization, up fatigue, balance, shrapnel damage caused errors.

5. The thermocouple do not suddenly pull out at high temperatures to prevent coat burst.

6. Keeping a clean furnace, removing oxides and other debris in the furnace.